This compatibility/stability test is a modified version of the ASTM D4740-94 method. The test methods list two separate procedures for predicting stability of residual fuel oil and the compatibility of residual fuel oil with a blend stock.
Part no.: 1001200 |
PRODUCT DESCRIPTION
Our density unit is fully in accordance with the specifications as mentioned in ASTM D1298 and IP 160.
The sample is poured into a density tube and this tube is then placed in a thermostatic controlled heater column to heat up the sample to 50ºC.
After checking the temperature a hydrometer is placed in the sample. Together with the hydrometer reading and a graph, one can now determine the density of the sample (fuel oil or lube oil) at 15ºC.
The density unit consists of the following equipment: - Density heater (thermostatic controlled heater column) - ASTM 12C thermometer - 4 hydrometers, range from 0,850 - 1.050 kg/cm3 - 2 density tubes - graph